- Direct reading on a calibrated tube
- Water: 0.1 – 1 l/h … 1.2 m³/h
- Air: 0.3 – 3 l/h … 16 m³/h
- Maxima: 10 bar or 100 °C
- Measuring glass tube: Conical (Pyrex®)
- Fittings: BSP ¼” or ½”
- Options :
- Pressure regulator
- Limit value switches
- Analogue output 4-20 mA
- ATEX version available
Fittings: | Threads or male or female accoding DIN EN ISO 228 T1 Option : Hose couplings |
Design: | From ”A” to ”Do” types Refer to datasheet (page 3) |
Operating temperature: | With NBR seals: Max. 80 °C With FPM seals: Max. 100 °C |
Operating pressure: | Max. 10 bar (Without water hammer event) |
Reading scale: | Engraved scale |
Graduated scale length: | SGK 1-2: ~150 mm SGK 3: ~220 mm |
Accuracy class: | VDE/VDI 3513, page 2 |
Error limit (G): | 1.6 % |
Linear limit (qG): | 50 % |
Calibration: | According application specifications |
MATERIALS
Body: | Anodized aluminum |
Fittings: | Standard: Or anodized aluminum, or PVC Options: AISI 316 Ti or PVDF |
Seals: | Standard: NBR for Aluminum version FPM for AISI 316 Ti version Options: EPDM, Perlast® (FFKM) |
Conical measuring tube: | Borosilicate glass |
Float: | Aluminum, PVC, PP, AISI 316 Ti Option: PTFE |
Pressure regulator: | Needle: AISI 316 Ti |
BAMO SGK 1-3 are used for direct reading and measuring of low flow rates on liquids and gases. They are designed for reduced space, in laboratories, in technical rooms.
The measuring principle is that of the variable area flowmeter, the float is moved by the fluid inside a calibrated conical glass tube. The glass tube (3) is sealed at each end by an O-ring protected against damage by a flat gasket (inserted in parts 1 and 2).
These flowmeters can be equipped with an adjustment valve and they receive limit value switches. The signal output from limit value switches is easily connected to a PLC or monitoring device. The GK2 version can be fitted with a ruler with a 4-20 mA transmitter (Hall effect sensors).
Standard versions:
- SGK-1: For smallest flows of air and water
- SGK-2: For medium flows of air and water
- SGK-3: Low flows of air and water
- Options:
- MSK1: With bistable Reed contact N.C.
- MSK2: With bistable Reed contact N.O.
- MSKW: With changeover Reed contact
- RC: ¹⁾ With inductive limit value switch
- EM: ²⁾ With analogue output 4-20 mA or 10 V
Note:
1) Flow-rates < 2 l/h of water; < 80 l/h of air
2) Available only with SGK-2 version (other ranges on request)
Limit value switches:
These flowmeters can be equipped with a magnetic float and Reed contacts, trigger points adjustable over the entire measuring range. For small flows up to 30 l/h, detection by inductive contact series RC will be used.
Analogue output 4-20 mA or 0-10 V:
The output signal is obtained by using a linear Hall effect displacement sensor. The output signal is proportional to the height of the magnetic float position. The sensor is plugged through a connector M12. This output is useful for applications using a PLC. The device is compact with a high level of measurement repeatability.