- 4 instruments DN10 to DN25
- Flow ranges from 0.3 up to 240 l/min
- Analogue output: 0/4-20 mA
- Digital output to set on site
- Back-lighted display
- Flow measurements on conductive and non-conductive liquids in various industrial, cosmetic and food applications. Examples: Demineralized water, liquid cosmetics, bases, acids, etc.
Diameter: | DN10, DN15, DN20, DN25 |
Range: | 0.3 to 21 L/min, 0.9 to 36 L/min, 3.5 to 60 L/min, 5.0 to 240 L/min |
Power supply: | 24V DC / 3.6W |
Outputs: | Digital output: to set as pulses (0.1 to 3000 ml/pulse) or empty tube alert, flowing back alarm, dosage, or min./max. flow alarm. Analogue output: 0/4 to 20 mA, scalable |
Display: | Alphanumeric back lighted LCD display |
Connections: | Connector M12 -5 pins (supplied) |
Accuracy: | ±2 % of reading and ±3 mm/s |
Repeatability: | ≤ 0.5 % |
Liquid temperature limits: | 0 to +80 °C |
Pressure limits: | 16 bar max. at 20 °C 10 bar max. at 20 °C |
Protection: | IP 67 |
Fittings: | BSP threads (standard) PVC Unions and other fittings on request: Straight tube, Clamp DIN 11864 |
Protection class: | IP 67 |
Materials: | PSU (Polysulfone); 2 EPDM flat seals |
Options: | |
Digital input: | To start a dosage (Connector M12 – 8 pins) |
Accuracy: | ±1 % of reading and ±3 mm/s, ±6 mm/s for NB 10 – 3/8” (according to VDI/VDE 2642) |
Remote interface: | To set all parameters (Adaptor USB/ RS485 with software), or remote display module |
Bamoflonic PSU applies propagation ultrasonic waves principle to measure the speed of a liquid, then to calculate the corresponding flow-rate. Therefore, It is convenient for liquids electrically conductive or not. Important: The liquid must be clear and homogeneous.
The absence of a moving part ensures its operation without mechanical wear.