Bamophar 107 pH/mV Monitor

Brand: BAMO
BAMO Mesures BAMOPHAR 107 pH-mV Wall Mounted Monitor, ISME Malaysia
  • Color touch screen
  • 2 existing scales: 0 to 14 pH or ±1000 mV
  • Temperature compensation: Automatic or manual
  • 2 outputs 0/4-20mA (pH and T °C)
  • 4 relays (Thresholds, alarm and/or regulation)
  • Options:
    • RS 422 /J-BUS + LOGGER
    • Extension terminal
  • Measurement of pH or ORP, alarms and/or regulation for water treatments, chemical industries, industrial applications
  • Examples:
    • Electroplating industry
    • Processed water treatment
    • Wastewater treatment (e.g. pH neutralization)
    • Groundwater or runoff water survey
    • Swimming pool pH regulation (or spa or fish tank)
    • Alarming on cooling plant (NH)
End-user interface:  Color touch screen 4.3”, resolution 480×272 pixels
Display of measurements, menus, temperature, relay status
Configuration – Keyword protected
Measuring scales:  0 to 14 pH as pH-meter or ± 1000 mV as mV-meter
Accuracy:  ±0.03 pH or ±3 mV
Input impedance:  >10¹³ Ω
Sensor signal input:  For coaxial connector (9054) on panel and rail mounting devices Screw connector on wall mounting devices
Temperature compensation: Automatic: Input for sensor Pt 100 Ω at 0 °C, range 0 to 100 °C
Manual: From 0 to 100 °C, by configuration
Relay outputs: 4 contacts N.O., voltage free
Configurable thresholds:  S1, independent threshold, to set up for measurement or temperature
S2, independent threshold, to set up for measurement or temperature
S3, independent threshold, to set up for measurement or temperature or external function
S4, threshold, to set up for alarming function
Too long injection – Timer exceeded
pH value out of range
Temperature sensor defect
or electrode cleaning mode
Contact initial resistance:  100 mΩ max. (voltage drop 6 V DC 1 A)
Switching power:  100 mA, 5 V DC ( 100 mA, 5 V DC (variable according to switching frequency, environmental conditions and accuracy).
ON/OFF regulation:  Adjustable cycle time from 0 to 9999 s, high and low proportional bandwidths, high and low dead zones
 PID regulation:  Adjustable proportionality from 0 to 200%, Integrant and Derivative: 0 to 999 s
 Calibration routine:  Relay outputs inhibited, Analogue outputs on standby at latest values
 Auto-cleaning menu:  Settings of frequency and duration, relay outputs inhibited, analogue outputs on standby at latest values
 Program testing:  Simulation through the menu on measurement, temperature, PID and relays output
 Measurement output:  0/4 – 20 mA (max. 600 Ω) proportional to measurement
 Temperature output/PID:  0/4-20 mA (max. 600 Ω), scaling 0 to 100 °C
This output is not available when PID regulation function is operating.
 Main power supply:  230 V AC / 50-60 Hz (others on request) – Consumption 10 VA
 Models:  Panel mounting, 96×144 mm, Front IP65, rear IP40
Wall mounting, IP65, cable glands
 Storage temperature:  -10 to +70 °C
 Operating température:  -5 to +50 °C
 Optional feature:  RS 422 + Logger
Interface: RS422 output, J-BUS link – Binary slave mode – 2400 to 9600 bauds
Data loggerRecord of cycle average measurement – 150 000 records max. on memory card.

Bamophar 107 pH/mV Monitor is equipped with a color touch screen for the display of a multilingual menu friendly and intuitive. It provides easy reading of measurement, temperature and state of the thresholds. It displays a menu with all parameters for configuration of analogue outputs, thresholds and regulation mode. In order to facilitate its commissioning, a programming menu can simulate the measurement, acting on the measurement analog outputs and P.I.D, as well as on the thresholds.

Analogue output reflects the measurement and may be scaled all along the range. Temperature analogue output is available as well as a 4-20 mA signal.

A complete measuring system setup with BAMOPHAR 107 pH/mV Monitor is as recommended below:

An extension terminal (wall, panel or DIN rail mounting):

  • Allows a second measuring parameter (pH, flow-rate, conductivity, etc.)
  • Data from this blind unit are displayed on the main unit
  • Connected to main unit with 4 wire shielded cable (Cable length between both devices: max. 500 m)
  • RS422 and Data Logger of main unit are shared between both units

Please contact us to learn more about the measuring system and the extension terminal.